BLOT-LINE EBV IgM

BLOT-LINE EBV IgM
EAN Code: 8595635307645
Catalog number: EBML20
Package size: 20 tests
Regulatory status: IVD CE
Storage: 2-8 °C
Producer: TestLine Clinical Diagnostics s.r.o.

Sample request

Sample ordering is possible only for registered users

If you are already registered, please log in
If you are new customer register

Contact person:

Name: MUDr. Jana Dočkalová
Phone: +420 549 121 205
E-mail: trade@testlinecd.com

The immunoblot assay is intended for the diagnosis of EBV infection using IgM antibodies to recombinant antigens in human serum or plasma in the general population. The qualitative and semi-quantitative manual assay is designed for professional use in a laboratory.

If you are already registered, please login

Request Product Export to PDF Print

Have a question?
Need a help?

Do not hesitate and call us or fill in a query form and we will contact you soon.

Description

Description:

·         Antigen used: combination of selected parts of the specific antigens (EBNA-1, VCA-p18, VCA-P23, EA-D).

·         Recombinant antigens are transferred to the nitrocellulose membrane (i.e. to the respective BLOTLINE (BL) strips).

·         If specific antibodies are present in the sample, they will bind to the respective antigens.

·         The complex is labelled with Conjugate and detected through a colour reaction with substrate (BCIP/NBT).

·         The kit enables 20 tests.

·         Short incubation periods, total assay time: approximately 1.5 h.

·         High sensitivity and specificity.

·         All reagents supplied are ready to use.

·         The kit may be used sequentially for smaller batches of samples (reagents are provided in sufficient quantities).

·         Positive and Negative controls included enabling validation of the tests.

·         BL strips equipped with conjugate control band and a control band indicating kits functionality and sensitivity.

·         Colour identification in compliance with antibody classes.

·         Identical (i.e. interchangeable) reagents (except for Conjugate) in all BLOT-LINE kits.

·         Evaluation of results – the intensity of the bands can be evaluated visually or by means of software (Immunoblot Software).

Applications:

·         Diagnostics of EBV.

·         Confirmatory test to ELISA.

Brief assay procedure:

1.       Dilute serum/plasma samples (1:51).

2.       Pipette diluted samples and controls into the channels of the tray and insert the BL strips.

3.       Incubate for 30 min at room temperature using a shaking apparatus.

4.       Wash 3 times for 5 min.

5.       Add Conjugate.

6.       Incubate for 30 min at room temperature using a shaking apparatus.

7.       Wash 3 times for 5 min.

8.       Add substrate (BCIP/NBT).

9.       Incubate for 15 min at room temperature using a shaking apparatus.

10.   Wash 2 times for 5 min in distilled water.

11.   Dry the strips and evaluate results using the validation strip enclosed or using Immunoblot Software.