Description:
·         Antigen used: combination of selected parts of the specific antigens of Yersinia spp. (YopB, YopD, YopE, YopH, YopM, YopN, LcrV, Ail, Invasin ) and  species-specific antigens of Y.enterocolitica (YstB, YscM-Y.Ent) and Y.pseudotuberculosis (YscM – Y.Pst).
·         Recombinant antigens are transferred to the nitrocellulose membrane (i.e. to the respective BLOT‑LINE (BL) strips).
·         If specific antibodies are present in the sample, they will bind to the respective antigens.
·         The complex is labelled with Conjugate and detected through a colour reaction with substrate (BCIP/NBT).
·         The kit enables 20 tests.
·         Short incubation periods, total assay time: approximately 1.5 h.
·         High sensitivity and specificity.
·         All reagents supplied are ready to use.
·         The kit may be used sequentially for smaller batches of samples (reagents are provided in sufficient quantities).
·         Positive and Negative controls included enabling validation of the tests.
·         BL strips equipped with conjugate control band and a control band indicating kits functionality and sensitivity.
·         Colour identification in compliance with antibody classes.
·         Identical (i.e. interchangeable) reagents (except for Conjugate) in all BLOT-LINE kits.
·         Evaluation of results – the intensity of the bands can be evaluated visually or by means of software (Immunoblot Software).
Applications:
·         Diagnostics of Yersinia.
·         Confirmatory test to ELISA.
Brief assay procedure:
1.       Dilute serum/plasma samples (1:51).
2.       Pipette diluted samples and controls into the channels of the tray and insert the BL strips.
3.       Incubate for 30 min at room temperature using a shaking apparatus.
4.       Wash 3 times for 5 min.
5.       Add Conjugate.
6.       Incubate for 30 min at room temperature using a shaking apparatus.
7.       Wash 3 times for 5 min.
8.       Add substrate (BCIP/NBT).
9.       Incubate for 15 min at room temperature using a shaking apparatus.
10.   Wash 2 times for 5 min in distilled water.
11.   Dry the strips and evaluate results using the validation strip enclosed or using Immunoblot Software.
 
					
					